Reliability of a portable device for quantifying tone and stiffness of quadriceps femoris and patellar tendon at different knee flexion angles.

Guoqian Chen | Jiatao Wu | Guocai Chen | Yanyan Lu | Wei Ren | Wu Xu | Xuemeng Xu | Zugui Wu | Yingxin Guan | Yi Zheng | Bofan Qiu
PloS one | 2019

The reliability of MyotonPRO that can monitor the mechanical properties of tissues is still unclear. This study aimed to analyze the within-day inter-operator and between-day intra-operator reliability of MyotonPRO for assessing tone and stiffness of quadriceps femoris and patellar tendon at different knee angles. The tone and stiffness of healthy participants (15 males and 15 females, aged 24.7±1.6 years) in the supine and resting position were measured using the MyotonPRO device. The measurements were quantified at 0°, 30°, 60°, and 90° of knee flexion. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were calculated and a Bland-Altman analysis was conducted to estimate reliability. The results indicated excellent inter-operator reliability (ICC > 0.78) and good to excellent intra-operator reliability (ICC > 0.41). The inter-operator SEM measurements ranged between 0.1-0.9 Hz and 3.8-37.9 N/m, and intra-operator SEM ranged between 0.5-1.3 Hz and 7.9-52.0 N/m. The inter-operator MDC ranged between 0.3-2.5 Hz and 10.5-105.1 N/m, and intra-operator SEM ranged between 1.1-3.3 Hz and 21.9-144.1 N/m. The agreement of inter-operator was better than that of intra-operator. The study concluded that MyotonPRO is a reliable device to detect the tone and stiffness of quadriceps femoris and patellar tendon.

Pubmed ID: 31365582

Research resources used in this publication

None found

Additional research tools detected in this publication

Antibodies used in this publication

None found

Associated grants

None

Publication data is provided by the National Library of Medicine ® and PubMed ®. Data is retrieved from PubMed ® on a weekly schedule. For terms and conditions see the National Library of Medicine Terms and Conditions.

This is a list of tools and resources that we have found mentioned in this publication.


MedCalc (tool)

RRID:SCR_015044

Statistical software for ROC curve analysis. MedCalc performs several statistical tests such as method comparison, method evaluation, reference intervals, and meta-analysis.

View all literature mentions