Suction devices for clearing the surgical field are among the most commonly used tools of every surgeon because a better view of the surgical field is essential. Forced suction may produce disturbingly loud noise, which acts as a nonnegligible stressor. Especially, in emergency situations with heavy bleeding, this loud noise has been described as an impeding factor in the medical decision-making process. In addition, there are reports of inner ear damage in patients due to suction noises during operations in the head area. These problems have not been solved yet. The purpose of this study was to analyse flow-dependent suction noise effects of different surgical suction tips. Furthermore, we developed design improvements to these devices.
Pubmed ID: 30584578
Publication data is provided by the National Library of Medicine ® and PubMed ®. Data is retrieved from PubMed ® on a weekly schedule. For terms and conditions see the National Library of Medicine Terms and Conditions.
Statistical analysis software that combines scientific graphing, comprehensive curve fitting (nonlinear regression), understandable statistics, and data organization. Designed for biological research applications in pharmacology, physiology, and other biological fields for data analysis, hypothesis testing, and modeling.